MF Series Thermal Mass Flow Meter and Controller: How to Choose Between MF30, MF50, MF51, MF52, MF60 and MF61

Table of Content [Hide]

    MF30, MF50, MF51 and MF61 thermal gas mass flow meters and controllers

    The MF series is not one flow meter with six options. It is six products that happen to share a sensor family, and they answer six different questions. Two of them are meters. Two of them are controllers with a valve inside. One is built for litres per hour, another for standard cubic centimetres per minute. Picking the wrong one is not a small mistake — it usually shows up as a range that never gets above 30% of scale, or a controller that cannot hold setpoint because the supply pressure is below its differential window.

    There is a second trap that catches even experienced buyers, and it is buried in a footnote: every flow range in the selection guide is calibrated on air. Measuring argon, propane or carbon dioxide on an air-calibrated meter does not give you the same span. The conversion table in the manual puts propane at 34.59% of the air range. A meter bought as a 1000 L/min air unit is a roughly 346 L/min propane unit.

    This guide covers what actually differs across MF30, MF50, MF51, MF52, MF60 and MF61 — the ranges, the accuracy classes, the units, the power requirements, the connectors and the gas conversion — so the model code you order matches the gas you are measuring.

    The Five Questions That Decide the Model

    Work through these before looking at a model number. Each one eliminates series quickly.

    QuestionWhat it eliminates
    What is the flow, in what unit?L/min, Nm³/h, NmL/min and SCCM/SLPM are four different families — not interchangeable
    Do you need to measure, or to control?MF60 and MF61 contain a normally-closed valve and accept a setpoint; the others only measure
    What accuracy does the application justify?±1.0% FS (MF60) versus ±(1.5+0.2FS)% (MF50, MF51) versus ±(2.0+0.2FS)% (MF30)
    What wetted material?Anodised aluminium, 304 or 316 body — a corrosive or high-purity gas removes F0
    Which gas, exactly?Determines the calibration digit in the model code and the usable span

    The unit question is the one that most often goes wrong in an enquiry. "Ten litres a minute" and "ten normal cubic metres an hour" are not the same order of magnitude, and a quotation cannot be compared until both sides are in the same unit and at the same reference condition.

    MF Series Comparison: MF30, MF50, MF51, MF52, MF60, MF61

    SeriesTypeFlow rangePipe / connectionAccuracyPowerResponseMax pressure
    MF30Meter, MEMS, LCDDN10 2.8–280 L/min
    DN15 6–600 L/min
    DN20 10–1000 L/min
    G3/8, G1/2, G3/4 external±(2.0+0.2FS)%DC12–24 V / 100 mA50–1000 ms≤1.6 MPa
    MF50Meter, MEMS, integral or splitDN8 1.8–180 L/min
    DN10 2.8–280
    DN15 6–600
    DN20 10–1000
    DN25 20–2000
    DN32 30–3000 L/min
    PT internal, DN8–DN32±(1.5+0.2FS)%DC12–24 V / 100 mA50–1000 ms≤1.6 MPa
    MF51Meter, platinum resistance sensor, stainless sleeveDN20 0.6–60 Nm³/h
    DN25 1–100
    DN32 1.6–160
    DN40 2.4–240 Nm³/h
    G3/4 to G1-1/2 internal±(1.5+0.2FS)%DC24 V / 1 A1–5 s (guide)
    1 s (manual)
    ≤1.6 MPa
    MF52Meter, micro flowS0 1–100 NmL/min
    S1 1–1000 NmL/min
    S2 1–100 NL/min
    S3 5–500 NL/min
    S4 10–1000 NL/min
    G1/4, PT1/2 internal; adapters<100 NL/min: ±(1.0+0.4FS)%
    >100 NL/min: ±(1.5+0.2FS)%
    DC12–24 V / 100 mA50 ms
    warm-up 3–4 min
    ≤1.0 MPa
    MF60Controller, normally-closed valveS0 1–100 SCCM
    … S8 3–300 SLPM
    G1/4, PT1/2 internalControl ±1.0% SP
    Measure ±1.0% FS (≤50 L/min), ±1.5% FS (>50 L/min)
    DC12–24 V / 6 WControl <1.5 s (T90)
    Measure <50 ms
    Measure 0–1.5 MPa
    MF61Controller, servo-motor valve, normally closedDN10 15–300 NL/min
    DN15 25–500
    DN20 50–1000
    DN25 100–2000
    DN32 150–3000 NL/min
    BSPT3/8 to BSPT1-1/4 internalControl ±1.0% SP
    Measure 1.5% FS (<1000 L/min)
    2.0% FS (>1000 L/min)
    DC12–24 V / 2.5 WControl <20 s (T90)
    Measure <200 ms
    Measure 0–1.5 MPa
    Control 0–1.0 MPa

    Two rows deserve a second look. MF51 draws 1 A at 24 V while MF30, MF50 and MF52 draw 100 mA at 12–24 V. If a panel supplies several of these from one small DC rail sized for the 100 mA family, MF51 will brown out. And MF51's response time is 1–5 seconds, not milliseconds — it is a thermal meter for larger gas lines, not a fast-response instrument.

    The Air Calibration Trap: Why the Same Meter Has a Different Range on Propane

    Every range in the selection guide is measured on air at 25 ℃ and 101.32 kPa. The manual states this twice per series, and then gives a conversion table for ten gases.

    GasCodeConversion factorRange as % of air figure
    AirA1.0100%
    OxygenB0.986198.61%
    NitrogenC0.99499.4%
    ArgonD1.4066140.66%
    Nitric oxideE0.970297.02%
    Nitrogen dioxideF0.736673.66%
    Carbon dioxideG0.732673.26%
    MethaneH0.714771.47%
    EthaneI0.478147.81%
    PropaneJ0.345934.59%

    Read the last column as the usable span on that gas relative to the air-calibrated figure. An MF50 DN20 is specified 10–1000 L/min on air. On propane, 34.59% of range means roughly 3.5–346 L/min. The instrument is not faulty and nothing needs recalibrating — but a line running at 600 L/min of propane will read over-range on a meter that looked comfortably sized on paper.

    Two consequences follow, and both belong in the enquiry.

    • Size on the working gas, not on air. Convert the process range through the factor first, then choose the model and size.
    • Order the gas code. The model code carries the medium: A air, B oxygen, C nitrogen, G carbon dioxide, MG mixed gas. A meter configured for air is not the same article as one configured for oxygen, even at the same size and range.

    On the MF30 the factor is also a menu parameter rather than only a factory setting: F12, gas correction factor, default 1000 meaning 1.0. That exists so a meter already in service can be moved to another gas — but it corrects the reading, it does not change the sensor's usable span. A meter rescaled in the menu to propane still has the same physical range it had on air.

    Argon is the useful counter-example. At 1.4066 it spans 140% of the air figure, so an argon line will read comfortably inside a meter that looked marginal on paper. Both directions of error are covered by the same table.

    Meter or Controller: MF60 and MF61 Are Different Products

    MF60 and MF61 are mass flow controllers. The manual lists a normally-closed valve, a control accuracy of ±1.0% SP, an input signal for the setpoint, and a response time quoted separately for control (<1.5 s, T90) and measurement (<50 ms). The others have none of that — they measure and output, and a separate valve or controller has to do the regulating.

    Choosing between the two controllers comes down to one line in the two manuals, and the two lines are almost opposite.

    MF60MF61
    Working differential pressure0.1–0.7 MPa (50 SLPM range); 0.1–0.6 MPa at 100 SLPM; 0.1–0.4 MPa at 300 SLPMNo pressure difference required
    Control response<1.5 s (T90)<20 s (T90)
    Turndown ratio1:1001:20
    PowerDC12–24 V / 6 WDC12–24 V / 2.5 W

    MF60 is a fast controller that lives on a regulated supply: it needs at least 0.1 MPa across the valve to have something to regulate against, and it will not hold setpoint on a 30 kPa house line however the electronics are configured. It rewards that requirement with a 1.5-second control response and a 1:100 turndown ratio.

    MF61 is the opposite trade. The manual states plainly that no pressure difference is required, so it can sit on a low-pressure line where MF60 cannot — at the cost of a 20-second control response and a 1:20 turndown ratio. For a slow purge, a blanketing flow or a steady process feed, that is entirely adequate. For leak testing or fast dosing, it is not.

    A mass flow controller installed on the wrong side of this table will either fail to hold setpoint or be far slower than the process expects. Both are piping and process decisions, and both need answering before the order.

    Both controllers also carry a stated IP40 rating and a selectable reference temperature — 0 ℃, 20 ℃ or 25 ℃, default 25 ℃. If the contract states normal conditions at 0 ℃, that digit has to be set at the factory, not converted later.

    Accuracy Is Not One Number

    The MF52 entry is the clearest example of why a single accuracy figure is misleading. The manual splits it at 100 NL/min:

    • Below 100 NL/min: ±(1.0+0.4FS)%
    • Above 100 NL/min: ±(1.5+0.2FS)%

    So the same series is a higher-accuracy instrument in its lower range and a standard one above it. If a quotation quotes one number for MF52 without saying which segment, the number is not comparable.

    The same logic explains why two suppliers quote different prices for what looks like the same meter at the same size. A DN20 at 10 L/min could be an MF30 at ±(2.0+0.2FS)% or an MF50 at ±(1.5+0.2FS)% — same pipe, same flow, different sensor family and different price. Accuracy class is a specification decision, not a marketing claim.

    Installation: Orientation, Straight Run and Valve Position

    The MF30 manual shows both horizontal and vertical installation as acceptable, and then sets the straight run by what sits upstream. The figures are in pipe diameters:

    Upstream conditionStraight run before the meter (a)Straight run after (b)
    Reducing or expanding pipe diameter20 × DN5 × DN
    One 90° elbow or tee20 × DN5 × DN
    Two 90° elbows in the same plane25 × DN5 × DN
    Two 90° elbows out of plane40 × DN5 × DN
    Control valve upstream55 × DN5 × DN

    The manual adds one instruction that is easy to lose in a P&ID: install the control valve and the buffer cut-off valve downstream of the meter, not upstream. A regulator immediately before the sensor puts the disturbance right where the straight run is supposed to be.

    These figures are from the MF30 manual. The proportion is the same physics across the series, but if the straight run is tight, check the multiplier in the manual for the model you are buying rather than applying 20 × DN by assumption.

    What the Meter Will Not Measure

    Three exclusions appear in the MF51 manual and apply to the family as a whole. They are worth filtering an enquiry on, because each one ends a project that looked fine on a flow range.

    • Acetylene is excluded. The manual specifies "all gases (except acetylene)" — the gas is outside the calibrated medium list and outside the safety envelope of a heated sensor.
    • The gas must be clean and free of particulate. The manual requires that the measured gas contain no particulate matter, because particles damage the sensor. A dirty line needs filtration before the meter, not a different meter.
    • A hazardous area needs the intrinsically safe version and a barrier. Where a flammable gas and air mixture is present, the manual requires an intrinsically safe meter used together with a certified safety barrier. That is a separate article number, not an option on the standard one.

    Wiring: Confirm the Connector Before You Cut Cable

    The same series can leave the factory with different connectors, and the pin assignments are not interchangeable between them. This is the one part of the specification that must be read off the terminal label rather than copied from a table.

    SeriesConnector documentedPin assignment and source
    MF304-core aviation cableMF30 manual §6: 1 red 24 V+, 2 black 24 V−, 3 white I+ / RS485 A, 4 green I− / RS485 B. The selection guide lists a 6-core plug instead — see the discrepancy section below.
    MF50M8 6-core, or Gx16 6-coreMF50 manual §6 gives two different pinouts. Gx16: 1 red 24 V+, 2 black 24 V− / I−, 3 green alarm output (PNP), 4 brown I+, 5 white RS485 B, 6 yellow RS485 A. M8: 1 pink 24 V− / I−, 2 black I+, 3 blue shell ground, 4 white 24 V+, 5 brown RS485 A, 6 grey RS485 B.
    MF516-core aviation plugSelection guide: 1 red 24 V+, 2 green pulse or PNP, 3 yellow RS485 A, 4 black 24 V− / I−, 5 brown I+, 6 white RS485 B.
    MF61DB9, 9-coreSelection guide: green supply +, blue supply −, yellow 4–20 mA I+, black 4–20 mA I−, red RS485 A, brown RS485 B.
    MF52Aviation socket; 4-pin micro USB on some sizes4-pin lead: red 24 V+, black 24 V−, white I+ / RS485 A, green I− / RS485 B.

    Where the Selection Guide and the Manuals Disagree

    The comparison table above follows the Chinese selection guide, because that is the document a quotation is built from. The individual English manuals do not always match it. These are the conflicts an engineer should settle before ordering, not after:

    ItemSelection guideSeries manual
    MF30 connector6-core aviation plug4-core aviation cable (§6)
    MF51 response time1–5 s1 s (§3.2)
    MF51 working pressure≤1.6 MPa≤3.0 MPa (§3.2)
    MF51 protectionnot statedIP64 (§3.2)
    MF51 accuracy±(1.5+0.2FS)%±1–2.5% (§3.2)
    MF51 storage temperature−10 to 65 ℃−40 to 80 ℃ (§2.2)
    MF51 DN20 rangetable 0.6–60 Nm³/hworked example 0.5–50 Nm³/h

    None of these is a rounding difference, and two of them change what the instrument can be used for: a 3.0 MPa rating instead of 1.6 MPa, and a sensor range of −40 to 250 ℃ instead of an ambient −10 to 55 ℃. The MF51 manual separates sensor and converter temperature — sensor −40 to +250 ℃, converter −20 to +55 ℃ — while the selection guide prints a single operating range.

    Where the two disagree, the manual shipped with the unit governs. For a quotation, ask which document the certificate and the nameplate follow. Publishing a comparison that silently mixes the two is how a meter arrives rated for 1.6 MPa on a 2.5 MPa line.

    Which Manual Covers Which Model

    The instruction manuals are separate documents per series, and the pin tables and dimension drawings differ between them. Match the document to the model code before wiring:

    SeriesManualKey content that differs
    MF30MF30 Series Instruction ManualG-thread connections, 6-pin aviation wiring, DN10–DN20 dimensions
    MF50MF50 Series Instruction ManualIntegral/split options, four display orientations, DN8–DN32 dimensions
    MF51MF51 Series Instruction ManualNm³/h ranges, 24 V/1 A supply, 1–5 s response, pulse output option
    MF52MF52 Series Instruction ManualNmL/min ranges, warm-up time, thread adapters (compression, quick-connect, hose barb)
    MF60MF60 Series Instruction ManualValve control, differential pressure window, IP40, selectable reference temperature
    MF61MF61 Series Instruction ManualBSPT connections, DB9 wiring, shared RS485 input/output

    Selection Checklist

    Send these seven items and the model code assembles itself:

    1. Gas and its code (A/B/C/G/MG), plus the actual composition if it is a mixture
    2. Normal flow and maximum flow, in the unit your process uses
    3. Reference conditions — 0 ℃, 20 ℃ or 25 ℃, and the pressure base
    4. Line pressure and, for a controller, the available differential pressure
    5. Required accuracy and whether the reading is for billing, control or indication
    6. Body material: anodised aluminium, 304 or 316
    7. Output and protocol: 4–20 mA, RS485 Modbus RTU, pulse, and whether both are needed simultaneously

    Frequently Asked Questions

    What is the difference between MF50 and MF30?

    Accuracy and size range. MF30 is ±(2.0+0.2FS)% with external G threads on DN10–DN20. MF50 is ±(1.5+0.2FS)%, covers DN8–DN32, and offers integral or split construction with four display orientations. For the same pipe size, MF50 is the tighter instrument.

    Can I use an MF50 on propane if it was calibrated on air?

    Yes, within the converted span. Propane's factor is 0.3459, so the usable range is about 34.6% of the air figure. Size the meter on the converted range, and order the gas code in the model number rather than relying on a field conversion.

    Do MF60 and MF61 measure, or do they also control?

    Both control. They contain a normally-closed valve and accept a setpoint through 4–20 mA, 1–5 V, 0–5 V, 0–10 V or Modbus, with a stated control accuracy of ±1.0% SP on MF60. MF30, MF50, MF51 and MF52 are meters only.

    Why does MF51 need 24 V at 1 A when the others use 100 mA?

    It is a different sensing construction for larger gas lines — a platinum resistance sensor in a stainless protective sleeve, with a 1–5 s response. Plan a dedicated supply rather than sharing a rail sized for the 100 mA family.

    Can I use 4–20 mA and RS485 at the same time?

    On the 4-pin lead, no — the manual states they are either/or. On the 6-core connectors (MF50, MF51) and the DB9 (MF61), both are wired separately and can be used together. MF30 depends on which connector is fitted — its own manual documents a 4-core cable, where the two are mutually exclusive.

    How long does MF52 need before the reading is valid?

    The manual gives a 3–4 minute warm-up. Readings taken immediately after power-up are not yet at the stated accuracy.

    Which series is IP rated?

    It depends on the document. The MF51 manual states IP64 and the MF60 and MF61 documents state IP40. The selection guide lists no protection rating for MF30, MF50 or MF52 — that is not the same as an IP40 rating, so for a washdown or outdoor location ask for the enclosure specification of the exact model rather than assuming.

    What is the default menu password?

    MF50 uses 1000, stated in its manual under the password entry section. MF30 uses 11111 — a five-digit default that can be changed in menu F91. The two are different, and an installer who has only seen one of the manuals will lock themselves out of the other.

    What reference temperature should I specify?

    Whatever the contract or the process data sheet uses. MF60 allows 0 ℃, 20 ℃ or 25 ℃ with 25 ℃ as the default, and the choice changes the numbers by several percent — the same reason normal and standard cubic metres are not interchangeable.

    Next Step

    Send the gas, the flow range in your unit, the reference condition and the line pressure. Those four determine the series, the size and the calibration digit, and they are also what turns a quotation into something you can compare against another supplier's.

    The thermal gas mass flow meter range covers the pipe-type, insertion and micro configurations built on the same sensor family. For oxygen service, where cleaning and material evidence matter more than range, see the oxygen flow meter selection guide.

    Ethan Liu
    Ethan Liu

    Ethan Liu is JUJEA's technical content specialist, translating complex sensor integration and industrial automation concepts into practical insights for engineers, system integrators, procurement professionals, and decision-makers worldwide.